Product Info

MetroLaser Standard Product

VibroMet™ 500 SLDV (Scanning Laser Doppler Vibrometer)
Specifications

The VibroMet™ 500 SLDV is based on a patented electro-optical configuration developed by MetroLaser. The system consists of a portable laser sensor head and electronic controller. The laser sensor head includes a laser, frequency modulator, photo-detector, color CCD camera, and two computer-controlled scanning mirrors. The electronic controller has both a demodulated velocity output and a 10.7 MHz frequency modulated signal output. The front of the controller has a signal strength indicator, two selectable velocity ranges and an array of low pass filter options. A comprehensive software package, including modal analysis, is available. The software registers the user-selected target, controls the mirrors, performs dithering if needed, collects and analyzes data, and displays 2D and 3D animations.

 

VELOCITY RANGE 5 µm/s to 800 mm/s
VIBRATION FREQUENCY RANGE 0.1 Hz to 40 kHz
WORKING DISTANCE 0.5 m to 5 m
OPTICS

Collimated
(No Focusing Needed)

SURFACE REFLECTIVITY ENHANCEMENT Typically None Required
SIGNAL OUTPUT Analog Demodulated and 10.7 MHz FM
DIMENSIONS - LASER HEAD 36 x 25 x 18 cm
DIMENSIONS - ELECTRONIC CONTROLLER 43 x 33 x 10 cm (standard 19-in rack)
WEIGHT - LASER HEAD 10 kg
WEIGHT - ELECTRONIC CONTROLLER 6.2 kg
LOW PASS FILTERS 1,2,5,10,20 kHz
OUTPUT VOLTAGE (max) +/ -8 Volts
LASER 780 nm, <15mW, Class IIIb
650 nm,  <1mW, Class II
SCAN RANGE +/- 20 degrees, each axis
BEAM SIZE 1 mm diameter
POWER REQUIREMENT 110V/220V