LARS · Qualification evidence · Experimental case study

Detecting designed internal voids
with LARS

An interactive engineering case study on exploring how nominal void fraction, build direction, similarity threshold, and analyzed spectral bandwidth influence defect detection in laser powder-bed-fused aluminum brackets.

AlSi10Mg · LPBF3 reference brackets3 spectral windowsThreshold 0.90–0.99Experimental / exploratory

Study objective and approach

Internal voids can alter a component's dynamic response. Laser Acoustic Resonance Spectroscopy (LARS) compares measured resonance signatures against reference measurements and assigns a similarity score. A score below the selected threshold flags a bracket for rejection or further review. This case study investigates how designed void fraction, defect location, build orientation, and analyzed spectral bandwidth influence detection performance, reference repeatability, and the tradeoff between defect detection and pristine acceptance.

Measured specimens and test conditions

Population Count Role Notes
Reference brackets 3 Reference construction L1, L3, L4; not specificity controls
Repeat measurements 8 Repeatability / expanded acceptance 3 × L1; 3 × L3; 2 × L4
Independent pristine 1Independent acceptance control Y7
Defective brackets 36 Observed detection fraction Large (M), medium (N), small (O), corner (P-V), side-Z (W-X)
Build-orientation specimens 4Separate sensitivity study Y1–Y4; excluded from confusion matrices

Bracket geometry & designed void locations

Bracket geometry and defect locations

Specimen identification

Specimen Defect / configuration Void fraction (%) [from CAD]

Reference and build-orientation specimens are identified separately from intentionally defective brackets.

Explore the results

All five plots, key metrics, and confusion matrices respond to the same analysis controls.

0.9000.9500.990

Five-plot evidence story

01 · VOID SENSITIVITY

Designed void fraction and observed defect detection

Nominal void fraction (%) is a designed-geometry proxy, not CT-measured porosity.

Defect groups at the same nominal void fraction may differ by location. Their detection probabilities are not interchangeable.
02 · OPERATING POINT

Threshold and frequency must be chosen together

Detection probability and pristine acceptance move together as the rejection threshold changes.

03 · MEASUREMENT STABILITY

Reference measurement repeatability

Reference and repeat measurements across three spectral windows.The dashed threshold line defines the minimum similarity required for acceptance. Repeat measurements below this line are flagged for rejection.

04 · SPECTRAL BANDWIDTH

Does a wider band help at low void fractions?

Paired specimen comparisons, with no assumption that individual modal mechanisms have been isolated.

The 10–30 → 10–60 comparison adds 30–60 kHz; 10–60 → 0–60 adds 0–10 kHz. These data do not test 60–80 kHz or identify specific vibration modes.
05 · MANUFACTURING VARIATION

Build-orientation sensitivity

Y1–Y4 are analyzed separately from defect POD and pristine specificity.

Decision summary

The table, selected operating point, and confusion matrix update when the controls change.

Key threshold results: 0.90, 0.95, 0.98

ThresholdFrequencyDetected defectsMissed defectsControls acceptedControls rejected

Confusion matrix at the selected operating point

Rows = actual condition; columns = LARS decision. The threshold rejects a bracket when its similarity score is strictly below T.

How to interpret these results

True positive (TP): a defective bracket correctly rejected. False negative (FN): a defective bracket incorrectly accepted. True negative (TN): a pristine control correctly accepted. False positive (FP): a pristine control incorrectly rejected.

POD = TP / (TP + FN); observed control acceptance = TN / (TN + FP). With only Y7 as an independent pristine control, statistical specificity for new production parts remains unestablished.

Experimental LARS case study · Nominal CAD void fraction · Toward production qualification · Empirical POD curves